Standard Practice for Using the Fricke Dosimetry System
Importancia y uso:
4.1 The Fricke dosimetry system provides a reliable means for measurement of absorbed dose to water, based on a process of oxidation of ferrous ions to ferric ions in acidic aqueous solution by ionizing radiation (ICRU 80, PIRS-0815, (4)). In situations not requiring traceability to national standards, this system can be used for absolute determination of absorbed dose without calibration, as the radiation chemical yield of ferric ions is well characterized (see Appendix X3).
4.2 The dosimeter is an air-saturated solution of ferrous sulfate or ferrous ammonium sulfate that indicates absorbed dose by an increase in optical absorbance at a specified wavelength. A temperature-controlled calibrated spectrophotometer is used to measure the absorbance.
Subcomité:
E61.02
Volúmen:
12.02
Número ICS:
17.240 (Radiation measurements)
Palabras clave:
absorbed dose; dosimetry; ferrous ammonium sulfate dosimeter; ferrous sulfate dosimeter; Fricke dosimeter; reference-standard dosimetry system;
$ 1,086
Norma
51026
Versión
23
Estatus
Active
Clasificación
Practice
Fecha aprobación
2023-12-01