Standard Test Methods for Static and Dynamic Characterization of Spinal Artificial Discs


Importancia y uso:

5.1 Artificial intervertebral discs are orthopaedic implants that replace degenerated natural intervertebral discs. Their function is to support the anterior column of the spine while allowing motion at the operated level. These test methods outline materials and methods for the characterization of the mechanical performance of different artificial intervertebral discs so that comparisons can be made between different designs.

5.2 These test methods are designed to quantify the static and dynamic characteristics of different designs of artificial intervertebral discs. These tests are conducted in vitro in order to allow for analysis of individual disc replacement devices and comparison of the mechanical performance of multiple artificial intervertebral disc designs in a standard model.

5.3 The loads applied to the artificial intervertebral discs may differ from the complex loading seen in vivo, and therefore, the results from these tests may not directly predict in vivo performance. The results, however, can be used to compare mechanical performance of different artificial intervertebral discs.

5.4 Fatigue tests should be conducted in a 0.9 % saline environmental bath at 37 °C at a rate of 2 Hz or less. Other test environments such as a simulated body fluid, a saline drip or mist, distilled water, or other type of lubrication could also be used with adequate justification. Likewise, alternative test frequencies may be used with adequate justification.

5.5 It is well known that the failure of materials is dependent upon stress, test frequency, surface treatments, and environmental factors. Therefore, when determining the effect of changing one of these parameters (for example, frequency, material, or environment), all others should be kept constant to facilitate interpretation of the results. In particular, it may be necessary to assess the influence of test frequency on device fracture while holding the test environment, implant materials and processing, and implant geometry constant.

Subcomité:

F04.25

Referida por:

F2789-25, F3292-25, F2694-26

Volúmen:

13.01

Número ICS:

11.040.40 (Implants for surgery, prothetics and orthotics)

Palabras clave:

artificial intervertebral disc; dynamic test methods; spinal implants; static test methods;