Standard Practice for Testing Variable-Wavelength Photometric Detectors Used in Liquid Chromatography


Importancia y uso:

4.1 Although it is possible to observe and measure each of the several characteristics of a detector under different and unique conditions, it is the intent of this practice that a complete set of detector specifications should be obtained under the same operating conditions. It should also be noted that to completely specify a detector’s capability, its performance should be measured at several sets of conditions within the useful range of the detector. The terms and tests described in this practice are sufficiently general that they may be used regardless of the ultimate operating parameters.

4.2 Linearity and response time of the recorder or other readout device used should be such that they do not distort or otherwise interfere with the performance of the detector. This requires adjusting the gain, damping, and calibration in accordance with the manufacturer's directions. If additional electronic filters or amplifiers are used between the detector and the final readout device, their characteristics should also first be established.

Subcomité:

E13.19

Referida por:

D6953-18

Volúmen:

03.06

Número ICS:

71.040.50 (Physicochemical methods of analysis)

Palabras clave:

linearity; noise measurement; photometric detector; variable wavelength; wavelength accuracy;