Standard Test Method for Mass Scale Calibration of Thermogravimetric Analyzers
Importancia y uso:
5.1 This test method calibrates or demonstrates conformity of thermogravimetric apparatus at ambient conditions. Most thermogravimetry analysis experiments are carried out under increasing temperature conditions or at isothermal temperatures distant from ambient conditions. This test method does not address the temperature effects on mass calibration.
5.2 In most thermogravimetry experiments, the mass change is reported as weight percent in which the observed mass at any time during the course of the experiment is divided by the original mass of the test specimen. This method of reporting results assumes that the mass scale of the apparatus is linear with increasing mass. In such cases, it may be necessary only to confirm the performance of the instrument by comparison to a suitable reference.
5.3 When the actual mass of the test specimen is recorded, the use of a calibration factor to correct the calibration of the apparatus may be required, on rare occasions.
Subcomité:
E37.01
Referida por:
E2958-21, E1868-10R21, E2550-21, E2402-24, E1131-20, E1131-20, E2403-23, E1582-21, E1131-20, E1131-20, E1641-23, C1872-24, E2008-17R21
Volúmen:
14.01
Número ICS:
19.020 (Test conditions and procedures in general), 71.040.50 (Physicochemical methods of analysis)
Palabras clave:
calibration; conformity; mass; thermogravimetry; thermogravimetric analyzer;
$ 1,053
Norma
E2040
Versión
25
Estatus
Active
Clasificación
Test Method
Fecha aprobación
2025-07-01