Standard Practice for Testing Nitrogen/Phosphorus Thermionic Ionization Detectors for Use In Gas Chromatography


Importancia y uso:

4.1 Although it is possible to observe and measure each of the several characteristics of a detector under different and unique conditions, it is the intent of this practice that a complete set of detector specifications be obtained at the same operating conditions, including geometry, flow rates, and temperatures. To specify a detector's capability completely, its performance should be measured at several sets of conditions within the useful range of the detector. The terms and tests described in this practice are sufficiently general so that they may be used under any chosen conditions.

4.2 Linearity and speed of response of the recorder should be such that it does not distort or otherwise interfere with the performance of the detector. Effective recorder response should be sufficiently fast so that its effect on the sensitivity of measurement is negligible. If additional amplifiers are used between the detector and the final readout device, their characteristics should first be established.

Subcomité:

E13.19

Referida por:

E0260-96R19

Volúmen:

03.06

Número ICS:

71.040.50 (Physicochemical methods of analysis)

Palabras clave:

flame ionization detector (FID); flame photometric detectors (FPD); gas chromatography (GC); nitrogen/phosphorus detectors (NPD); packed columns; supercritical fluid chromatography (SFC); thermionic detectors;