Standard Test Method for Retention Characteristics of 0.40 to 0.45 μm Membrane Filters Used in Routine Filtration Procedures for the Evaluation of Microbiological Water Quality 


Importancia y uso:

5.1 Microbiological water testing procedures using membrane filtration are based on the premise that all bacteria within a specific size range will be retained by the membrane filter used. If the membrane filter does not retain these bacteria, false negative results or lowered density estimates may occur that could have serious repercussions due to the presence of unrecognized potential health hazards in the water being tested, especially in drinking water.

5.2 This procedure as devised will enable the user to test each membrane filter lot number for its ability to retain all bacteria equal to, or larger than, the stated membrane pore size.

Subcomité:

D19.24

Volúmen:

11.02

Número ICS:

07.100.20 (Microbiology of water), 13.060.70 (Examination of biological properties of water)

Palabras clave:

filter; membrane; microbiological; water quality ;