Standard Practice for Noise Equivalent Temperature Difference of Thermal Imaging Systems
Importancia y uso:
5.1 This practice gives an objective measure of the temperature sensitivity of a thermal imaging system (relative to a standard reference filter) exclusive of a monitor, with emphasis on the detector(s) and preamplifier.
Note 1: Test values obtained under idealized laboratory conditions may or may not correlate directly with service performance.
5.2 This practice affords a convenient means for periodically monitoring the performance of a given thermal imaging system.
5.3 NETD relates to minimum resolvable temperature difference as described in Practice E1213. Thus, an increase in NETD may be manifest as a loss of detail in imagery.
5.4 Intercomparisons based solely on NETD figures may be misleading.
Note 2: NETD depends on various factors such as spectral bandwidth and background temperature.
Subcomité:
E07.10
Referida por:
E2533-21, E3353-22
Volúmen:
03.03
Número ICS:
17.200.20 (Temperature-measuring instruments)
Palabras clave:
infrared imaging systems; noise equivalent temperature difference; nondestructive testing; thermal imaging systems; thermography;
$ 949
Norma
E1543
Versión
14(2022)
Estatus
Active
Clasificación
Practice
Fecha aprobación
2022-12-01