Standard Practice for Use of a Calibration Device to Demonstrate the Inspection Capability of an Interferometric Laser Imaging Nondestructive Tire Inspection System
Importancia y uso:
5.1 All S/H systems change with time and use. Therefore, a calibration procedure for evaluating the operation of an S/H system is desirable. This calibration procedure provides a method of obtaining an optimized interferometric image pattern associated with a given size anomaly.
5.2 The use of straining blocks as calibration devices provides a means for ensuring the continued optimal performance of the S/H system. Straining blocks can also be used to compare performance of S/H systems in different facilities.
5.3 At not greater than a three (3) month interval the S/H system shall be calibrated following the procedures described in this practice. When necessary, adjustments, repairs, or modifications shall be made to the S/H system until it is able to observe, in the same image, all anomalies of size within the range of interest contained in the straining blocks.
Subcomité:
F09.10
Referida por:
E2581-14R23, E2533-21
Volúmen:
09.02
Número ICS:
31.260 (Optoelectronics. Laser equipment)
Palabras clave:
calibration; holography; inspection; nondestructive; shearography; tire;