Standard Test Method for pH of Atmospheric Wet Deposition Samples by Electrometric Determination


Importancia y uso:

5.1 The accurate measurement of pH in atmospheric wet deposition is an essential and critically important component in the monitoring of atmospheric wet deposition for trends in the acidity and overall air quality. Atmospheric wet deposition is, in general, a low ionic strength, unbuffered solution. Special precautions, as detailed in this test method, are necessary to ensure accurate pH measurements (1).3 Special emphasis must be placed on minimizing the effect of the residual liquid junction potential bias.

5.2 This test method is applicable only to the measurement of pH in atmospheric wet deposition. Its use in other applications may result in inaccuracies.

5.3 Fig. 1 provides a frequency distribution of precipitation pH values measured in conjunction with a national monitoring program within the United States. These data are an indication of the range of pH values common to atmospheric wet deposition.

FIG. 1 Frequency Distribution of Measured Laboratory pH of Atmospheric Wet Deposition From the 1984 National Atmospheric Deposition Program (NADP)/National Trends Network (NTN)


Subcomité:

D22.03.02

Referida por:

D1356-20A, D1356-20A, D1356-20A, D6328-18

Volúmen:

11.07

Número ICS:

71.040.50 (Physicochemical methods of analysis)

Palabras clave:

atmospheric deposition; hydrogen ion; low ionic strength solutions; pH; precipitation; wet deposition;