Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
Importancia y uso:
5.1 This test method is one of those required to determine if the presence of a medical device may cause injury to individuals during an MR examination or in the MR environment. Other safety issues which should be addressed include, but may not be limited to: magnetically induced torque (see Test Method F2213) and radiofrequency (RF) heating (see Test Method F2182). The terms and icons in Practice F2503 should be used to mark the device for safety in the magnetic resonance environment.
5.2 If the maximum magnetically induced displacement force for the specified magnetic field conditions (see Appendix X3) is less than the force on the device due to gravity (its weight), it is assumed that any risk imposed by the application of the magnetically induced force is no greater than any risk imposed by normal daily activity in the Earth’s gravitational field. This statement does not constitute an acceptance criterion; it is provided as a conservative reference point. It is possible that a greater magnetically induced displacement force can be acceptable and would not harm a patient or other individual in a specific case.
Note 2: For instance, in the case of an implanted device that is or could be subjected to a magnetic displacement force greater than the force due to gravity, the location of the implant, surrounding tissue properties, and means of fixation within the body may be considered. For a non-implanted device with a magnetically induced force greater than the gravitational force, consideration should be given to mitigate the projectile risk which may include fixing or tethering the device or excluding it from the MR environment so that it does not become a projectile.
5.3 The maximum static magnetic field strength and spatial field gradient vary for different MR systems. Appendix X3 provides guidance for calculating the allowable static magnetic field strength and spatial field gradient.
5.4 This test method alone is not sufficient for determining if a device is safe in the MR environment.
Subcomité:
F04.15
Referida por:
F1831-17, F2182-19E02, F3160-21, F3395_F3395M-19, F2119-24, F2213-17, F2503-23E01
Volúmen:
13.01
Número ICS:
11.040.40 (Implants for surgery, prothetics and orthotics)
Palabras clave:
medical device; metals (for surgical implants and medical devices); MRI (magnetic resonance imaging); MR safety;
$ 1,122
Norma
F2052
Versión
21
Estatus
Active
Clasificación
Test Method
Fecha aprobación
2021-10-01
