Standard Test Method for On-Line Measurement of Low-Level Dissolved Oxygen in Water


Importancia y uso:

5.1 DO may be either a corrosive or passivating agent in boiler/steam cycles and is therefore controlled to specific concentrations that are low relative to environmental and wastewater treatment samples. Out-of-specification DO concentrations may cause corrosion in boiler systems, which leads to corrosion fatigue and corrosion products — all detrimental to the life and efficient operation of a steam generator. The efficiency of DO removal from boiler feedwater by mechanical or chemical means, or both, may be monitored by continuously measuring the DO concentration before and after the removal process with on-line instrumentation. DO measurement is also a check for air leakage into the boiler water cycle.

5.2 Feedwater chemistry guidelines for high-pressure boilers generally require specific feedwater DO concentrations: 5 μg/L or less for reducing all volatile treatment [AVT(R)]; 5–10 μg/L for oxidizing all volatile treatment [AVT(O)]; 50–200 μg/L for oxygenated treatment [OT] (3).

5.3 In microelectronics production, DO can be detrimental in some manufacturing processes, for example, causing undesirable oxidation on silicon wafers.

Subcomité:

D19.03

Referida por:

D5127-13R18

Volúmen:

11.02

Número ICS:

13.060.50 (Examination of water for chemical substances)

Palabras clave:

dissolved oxygen; galvanic cell; oxygen; polarographic sensor;