Standard Test Method for Force Calibration of Thermomechanical Analyzers


Importancia y uso:

5.1 Most thermomechanical analysis experiments are carried out with some force applied to the test specimen. This force is often created electronically. It may be constant or changed during the experiment.

5.2 This method demonstrates conformance or calibrates the electronically applied force signal.

5.3 This method may be used for research and development, quality control, manufacturing or regulatory applications.

5.4 Other thermomechanical analyzer calibration functions include temperature by Test Method E1363 and length change by Test Method E2113.

Subcomité:

E37.10

Referida por:

E2347-21, E2092-23, E2769-22

Volúmen:

14.01

Número ICS:

71.040.50 (Physicochemical methods of analysis)

Palabras clave:

calibration; conformity; force; thermal analysis; thermomechanical analysis;

$ 949

Agregar al carrito

Norma
E2206

Versión
21

Estatus
Active

Clasificación
Test Method

Fecha aprobación
2021-03-01