Standard Practice for Evaluating Mobile Bearing Knee Tibial Baseplate Rotational Stops
Importancia y uso:
4.1 Fundamental aspects of this practice include the use of dynamic rotational force and motion representative of the human knee joint during an activity of daily living (deep flexion) and the effect of these forces and motions on the design features which stop or limit rotation in a mobile bearing knee design.
4.2 This test is required if rotational stops are designed to limit motion to ±20° or less; or there are other resistances to rotational motion with this ±20° range. In some instances, the rotational displacement could occur in both the inferior and superior interfaces.
FIG. 1 Schematic of Test Setup
Subcomité:
F04.22
Volúmen:
13.02
Número ICS:
11.040.40 (Implants for surgery, prothetics and orthotics)
$ 978
Norma
F2722
Versión
21
Estatus
Active
Clasificación
Practice
Fecha aprobación
2021-02-15