Standard Practice for Removal of Uranium or Plutonium, or both, for Impurity Assay in Uranium or Plutonium Materials
Importancia y uso:
5.1 This practice can be used to separate uranium or plutonium, or both, prior to the impurity analysis by various techniques. The removal of uranium and plutonium prior to quantification can improve the detection limits by minimizing the signal suppression caused by uranium or plutonium when using ICP techniques. Detection limits of ~1–10 part-per-billion (PPB) may be obtainable by matrix removal. Also, removal of the uranium and plutonium may allow the impurities analysis to be performed on a non-glove box enclosed instrument.
5.2 Other test methods exist to determine impurities in uranium or plutonium. Test Method C1517 is able to determine many impurities in uranium at detection levels of ~1–10 part-per-million (ppm) by DC-Arc Spectrometry. Test Method C1287 is able to determine impurities in uranium at detection levels of ~100 ppb by ICP-MS. Test Method C1432 provides an alternative technique to remove plutonium by ion exchange prior to analysis of the impurities by ICP-AES.
5.3 This practice can be used to demonstrate compliance with nuclear fuel specifications, for example, Specifications C753, C757, C776, C787, C788, and C996.
Subcomité:
C26.05
Volúmen:
12.01
Número ICS:
71.040.50 (Physicochemical methods of analysis)
Palabras clave:
AAS; extraction chromatography; ICP-AES; ICP-MS; impurities; plutonium; resin; separation; uranium;