Standard Practice for Subcutaneous Screening Test for Implant Materials


Importancia y uso:

4.1 This practice is a guideline for a short-term screening test for the evaluation of the tissue response to materials that may be selected for implantation in the human body and should be done in accordance with good laboratory practices. This test may be performed prior to long-term testing (for example, Practice F981) to eliminate unsuitable candidate materials early and to avoid unnecessary animal testing.

4.2 This practice may be used to detect toxic effects of materials in general (see Appendix X1). However, it is particularly suitable for the testing of materials that are intended to have contact with subcutaneous tissues or soft tissues in general. For materials intended to be inserted specifically into muscle tissues, Practice F763 should be considered as a short-term test method.

4.3 The suggested implant specimens are cylindrical. A special grooved type of cylinder may be used (see Fig. X2.1 of Appendix X2) to allow tissue interlocking that could keep the implant in place and minimize tissue irritation through motion at the interface that otherwise could contribute to increased variance of the results. In case ungrooved cylinders are used (see Fig. X1.2 of Appendix X2), probable motion at the implant/tissue interface must be taken into account. Control cylinders should be shaped like the test cylinders.

4.4 The type of surface preparation of the specimens can affect the tissue reaction; therefore the preparation procedure should be noted in the report. The test may be used to compare the effect of different surface structures or conditions of the same material or to assess the effect of various treatments of modifications of a material.

Note 1: If this method is used for material research, testing for endotoxin prior to implantation should be considered.

Subcomité:

F04.16

Referida por:

F0748-16, F2066-23, F1983-23, E3219-20

Volúmen:

13.01

Número ICS:

11.040.40 (Implants for surgery, prothetics and orthotics)

Palabras clave:

biocompatibility; medical devices; mice; short-term tissue screening; subcutaneous tissue screening; tissue compatibility; toxicity/toxicology ;

$ 1,122

Agregar al carrito

Norma
F1408

Versión
20a

Estatus
Active

Clasificación
Practice

Fecha aprobación
2020-08-01