Standard Test Method for Analysis of Target Phenols (TPs) in Soil by Multiple Reaction Monitoring Liquid Chromatography/Mass Spectrometry (LC/MS/MS)


Importancia y uso:

5.1 This test method developed for the analysis of TPs in soil and sediment samples is based upon an LC/MS/MS analysis. Any type of coupled liquid chromatography/mass spectrometry system may be used that meets the study objectives of the individual project. These may include, but are not limited to: trap, single quadrupoles, time-of-flight, high resolution, and others not mentioned here.

5.2 The MDL and reporting range for TPs are listed in Table 1. This SOP has been tested on Ottawa sand, four ASTM soil types, biosolid sample, and one commercial soil. The P&A QC acceptance criteria are listed in Table 3. Tables 4-17 display the TC and surrogate recoveries in the various soil types. 40 CFR Part 136, Appendix B was used as a guide to determine the MDLs. The 40 CFR Part 136 MDL criteria were not met for NP2EO; this does not affect the method because the SOP only reports to the RL and is not a regulatory method. All site sample results are not reported below the RL using this method. RLCS concentrations may be reported below the RL because they are spiked at or near the RL.

(A) Uncertainty calculation based upon 95 % confidence interval and a two-tailed Student t distribution.

   Uncertainty = Student t Value [(standard deviation) / (number of LCS)1/2]. (A) P&A values are after subtraction of average of MB if ≥RL. (A) P&A values are after subtraction of average of MB if ≥RL. (A) <RL = below reporting limit.

5.3 The RL for a specific soil sample may differ from that listed depending on the nature of the interferences in the sample matrix. Variability in historical LCS spike recovery may be used to estimate uncertainty. The estimate of minimum laboratory contribution to measurement uncertainty of this test method for each analyte is listed in Table 3. These values are derived from P&A samples from the initial IDOC study for this test method. The uncertainty will be greater near the RL and much greater near the DL. Also, uncertainty estimated based on variability in LCS recovery is conservative because some sources of variability are not included, such as subsample variability and matrix analyte recovery. This SOP covers multiple soil matrices and the uncertainty among the various matrices is variable.

Subcomité:

D34.01.06

Volúmen:

11.04

Número ICS:

13.080.10 (Chemical characteristics of soils)

Palabras clave:

Bisphenol A; LC/MS/MS; liquid chromatography; mass spectrometry; nonylphenol; octylphenol; sediment; soil;