Standard Test Method for Determining Particle Size Distribution of Alumina or Quartz by Laser Light Scattering


Importancia y uso:

4.1 It is important to recognize that the results obtained by this method or any other method for particle size distribution utilizing different physical principles may disagree. The results are strongly influenced by the physical principles employed by each method of particle size analysis. The results of any particle sizing method should be used only in a relative sense, and should not be regarded as absolute when comparing results obtained by other methods.

4.2 Light scattering theory that is used for determination of particle size has been available for many years. Several manufacturers of testing equipment have units based on these principles. Although each type of testing equipment utilizes the same basic principles for light scattering as a function of particle size, different assumptions pertinent to applications of the theory and different models for converting light measurements to particle size may lead to different results for each instrument. Therefore, the use of this test method cannot guarantee directly comparable results from the various manufacturers' instruments.

4.3 Manufacturers and purchasers of alumina and quartz will find the method useful to determine particle size distributions for materials specifications, manufacturing control, and research and development.

Subcomité:

C28.03

Referida por:

D5377-93R21, F1538-24

Volúmen:

15.01

Número ICS:

81.060.10 (Raw materials)

$ 946

Agregar al carrito

Norma
C1070

Versión
01(2020)

Estatus
Active

Clasificación
Test Method

Fecha aprobación
2020-01-01